Skip to main content
. 2016 Sep 7;6(3):39. doi: 10.3390/brainsci6030039

Table 3.

Maximum spatial extension (mm) of the 0.2 V/mm electric field isolevel achieved at each plane for voltage (3 V) and current (3.4 mA) controlled stimulation for all leads. Measurements performed at the ZI.

Plane 3389 6148 6180 SureStim1
Voltage Current Voltage Current Voltage Current Voltage Current
AXIAL 3.34 3.85 3.46 3.84 3.29 3.86 3.23 3.94
SAGITTAL 3.40 3.87 3.50 3.85 3.35 3.90 3.17 3.90
CORONAL 3.50 3.83 3.55 3.80 3.32 3.84 3.23 3.88