RCmean and RCmax ranges as defined by EARL [8], compared with RC results for all spheres at 10 min and 62 s scan duration

Sphere volume (mL) EARL: RCmean range RCmean at t = 600 s RCmean at t = 62 s EARL: RCmax range RCmax at t = 600 s RCmax at t = 62 s
26.52 0.76–0.89 0.79 0.80 0.95–1.16 0.98 1.05
11.49 0.72–0.85 0.75 0.74 0.91–1.13 0.96 1.04
5.57 0.63–0.78 0.72 0.69 0.83–1.09 0.97 0.94
2.57 0.57–0.73 0.68 0.64 0.73–1.01 0.93 0.90
1.15 0.44–0.60 0.44 0.48 0.59–0.85 0.59 0.71
0.52 0.27–0.38 0.33 0.27 0.31–0.49 0.44 0.40

For all spheres, RCs were within EARL specifications