Table 3. tested and recommended products for working with Qdots.

Original Vivid Notes
PBS Quenching Quenching avoid (even if manufacturer’s protocol still recommends all solutions in PBS)
TBS photobleaching Quenching & photobleaching avoid
Bond Wash photobleaching Quenching & photobleaching avoid
Tween @ 0.1% quenching quenching “stable up to 0.5% “in manufacturer information
avoid
BSA Provides some protection from quenching (in Tris) Test further if needed
Qdot diluant Good—Moderate quenching in solution Used in multiplex
Borate buffer Moderate quenching in solution Use with caution
TRIS quenching Use with caution to dilute BKRA if cost needs to be kept low
BKRA stable Used in multiplex
SFB Fairly stable Used in multiplex
Qdots
525 Stable Original Qdot–very stable and reliable in most buffer tested
565 Stable Original Qdot–very stable and reliable in most buffer tested
585V Least stable in all buffer tested including BKRA Avoid if possible Control for stability at all times
605V Average stability Control for stability at all times
625V Good stability Control for stability
655V Average stability Control for stability at all times
705V Good stability overall Control for stability
Cytoseal Good stability for 1 to 2 weeks Control for loss of fluorescence with time if quantitative analysis over more than 1 day
Ecomount Good stability for at least 2 weeks recommended