Table 1. Mean general trueness and standard deviation (SD), in μm, for partially and totally edentulous maxilla, and p values testing the scanner by model interaction.

N = 5 scans for each scanner and model type.

Scanner Partially edentulous maxilla Totally edentulous maxilla p-value1
Mean Trueness ± SD (μm) Mean Trueness ± SD (μm)
Trios® 71.2 ± 19.5 † 71.6 ± 26.7 † 0.9
CS 3500® 47.8 ± 7.3 ‡ § 63.2 ± 7.5 ‡ § 0.4
Zfx Intrascan® 117.0 ± 28.6 ‡ ¶ 103.0 ± 26.9 ‡ ¶ 0.5
Planscan® 233.4 ± 62.6 † § ¶ 253.4 ± 13.6 † § ¶ 0.3

The same symbol after SD indicates differences in trueness between scanner pairs (Tukey-adjustment for multiple comparison). Minimum significant difference across scanners: 65.1 μm and 37.1 μm for partially and totally edentulous maxilla models, respectively.

1 p-value testing of the interaction between scanner and model type (partially vs totally edentulous maxilla): a p-value > 0.05 indicates no difference in scanner trueness according to model type.