Table 1.

Acceptable plan criteria for prostate IMRT.

PTV CTV
Ideal D95 > 81 Gy
D99 > 76.95 Gy
D100 = 81 Gy
Mean Dose
81–84.24 Gy
Minor violation (acceptable) D90 > 81 Gy
or
D98 > 76.95 Gy
D99.0 > 81 Gy
Mean Dose
84.24–85.46 Gy

Minor violation: allowable when organs at risk constraint cannot be met. Dx refers to the dose that covers X percent of the structure. The prescribed dose was 81 Gy. Gy: Grey.