Skip to main content
. Author manuscript; available in PMC: 2017 Nov 1.
Published in final edited form as: Ultrasound Med Biol. 2016 Aug 9;42(11):2662–2675. doi: 10.1016/j.ultrasmedbio.2016.06.018

TABLE 2.

Thresholds of P +(aVg) and Eeff, and slope of regression curve for SC vs. ln(P+(avg)) or SC vs. Eeff after 1,000 shocks in water and 1,3-Butanediol.

Medium in
the holder
Size
(mm)
Threshold
P+(aVg)(MPa)/Eeff(mJ)
Slope
SC vs. ln(P+(avg))Eeff
Water 4 12.7 (12.3, 13.1)/
20.1 (18.4, 22.5)
0.88(0.75, 1.02)/
0.019 (0.015,0.023)
7 8.8 (8.7,9.1)/
7.9 (7.2,8.8)
0.82 (0.77,0.88)/
0.021 (0.019,0.022)
10 7.9 (7.5,8.3)/
3.8 (2.1,5.6)
0.63 (0.58,0.68)/
0.016 (0.015,0.018)
1,3-Butanediol 4
7 9.5 (8.3,11.0)/
6.8 (-0.2,14.9)
0.25(0.14,0.36)/
0.005 (0.002,0.008)
10 9.1 (8.4,9.8)/
7.8 (3.7,11.2)
0.30 (0.22,0.38)/
0.007 (0.005,0.009)