Skip to main content
. 2016 Oct 4;214(Suppl 3):S192–S202. doi: 10.1093/infdis/jiw150

Table 2.

Initial LoD Testing on Ebola Virus (EBOV) Strain Makona and Sudan Virus (SUDV) Strain Gulu Spiked in EDTA Whole Blood Samples, Using Reference RT-PCR Assays

Strain, Dilution in PFU/mL Theoretical Genomic Concentration, Copies/mL Experimental Genomic Concentration, Copies/mL
Detection Rate, Proportion of Samples
Detection Rate, Idylla™ Prototype Ebola Virus Test, Proportion of Samples
EZ1 RT-PCR
EBOV strain Makona
 Undiluted stock 81 500 000 81 500 000 3/3 3/3
 10 000 27 167 69 547 3/3 3/3
 1000 2717 3837 3/3 3/3
 100 272 135 3/3 3/3a
 10 27 ND 0/3 1/3a
 1 3 ND 0/3 0/3
Trombley RT-PCR
SUDV strain Gulu
 Undiluted stock 182 000 000 182 000 000 3/3 3/3
 10 000 206 116 134 818 3/3 3/3
 1000 20 612 26 902 3/3 3/3
 100 2061 2838 3/3 3/3a
 10 206 812 2/3 1/3
 1 20 122 1/3 0/3

Abbreviations: EBOV, Ebola virus; GP, glycoprotein; ITM, Institute of Tropical Medicine; LoD, limit of detection; ND, not detected; PFU, plaque-forming units; RT-PCR, reverse-transcription polymerase chain reaction; SUDV, Sudan virus.