Summary of Clinical Performance Specificity and Sensitivity

Characteristic, Specimen Negative Positive Negative Agreement
Samples, Proportion Percentage (95% CI)
Specificity
 Normal serum panels 131 7 131/138 94.9 (89.8–97.9)
 Finger stick 39 1 39/40 97.5 (86.8–99.9)
 Whole-blood replicates 226 8 226/234 96.6 (93.4–98.5)
 Serum replicates 200 49 200/249 80.3 (93.4–98.5)
Positive Agreement
Samples, Proportion Percentage (95% CI)
Spiked sensitivity, ng/test
 Whole blood
  9.4 46 71 71/117 60.7 (51.2–69.6)
  18.8 2 90 90/92 97.8 (92.2–99.7)
 Serum
  4.7 16 123 107/123 92.9 (79.8–91.9)
  9.4 6 93 93/99 93.9 (87.2–97.4)