Table 2.

Summary of available wear testing data for cervical disc replacements under idealized (Mode I) conditions

Reference Design Bearing couple Standard referenced Relevant test inputs Wear rate Particle size
SSED for PMA Number: P060023 (Bryan® Cervical Disc) [21] Bryan® (Medtronic Sofamor Danek, Memphis, TN, USA) Polycarbonate urethane on titanium N/A FE: ± 4.9°
LB: N/A
AR: ± 3.8°
AL: 130 N and 300 N
F: 2, 4, and 6 Hz
As published:
0.4–1.4 mg/ MC
Converted (density = 1.2 mg/mm3): 0.3–1.2 mm3/MC
Greater than 90% of particles were less than 1 µm
SSED for PMA Number: P070001 (ProDisc™-C Total Disc Replacement) [20] ProDisc™-C (Synthes Spine, West Chester, PA, USA) UHMWPE on CoCr N/A FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 150 N
F: 1.0 Hz
As published:
2.59 ± 0.36 mg/MC
Converted (density = 0.94 mg/mm3): 2.8 ± 0.36 mm3/MC
Average size 0.17–0.35 µm
Nectow et al., 2008 [43] ProDisc™-C (Synthes Spine) UHMWPE on CoCr ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.99 ± 0.15 mg/MC
Converted (density = 0.94 mg/mm3): 2.12 ± 0.15 mm3/MC
Greater than 90% of particles were less than 1 µm
Bushelow et al., 2008 [12] ProDisc™-C (Synthes Spine) UHMWPE on CoCr N/A FE: ± 5°
LB: ± 5.0°
AR: ± 3.5°
AL: 15-150 N
F: 1.0 Hz
As published:
0.88 ± 0.07 mg/MC
Converted (density = 0.94 mg/mm3): 0.94 ± 0.07 mm3/MC
Greater than 90% of particles were less than 1 µm; average diameter between 0.22 and 0.37 µm
N/A FE: ± 10°
LB: ± 5.0°
AR: ± 4.0°
AL:150 N
F: 1.0 Hz
As published:
2.59 ± 0.36 mg/MC
Converted (density = 0.94 mg/mm3): 2.75 ± 0.38 mm3/MC
ISO 18192-1 FE: ± 7.5°
LB: ± 4.0°
AR: ± 6.0°
AL: 15-150 N
F: 1.0 Hz
As published:
1.82 ± 0.11 mg/MC
Converted (density = 0.94 mg/ mm3): 1.94 ± 0.12 mm3/MC
SSED for PMA Number: P110002 (Mobi-C® Cervical Disc Prosthesis) [24] Mobi-C® (LDR Spine, Austin, TX, USA) UHMWPE on CoCr ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.55 ± 0.08 mg/MC
Converted (density = 0.94 mg/mm3): 1.64 ± 0.09 mm3/MC
Average size 0.77 µm
SSED for PMA Number: P100003 (SECURE®-C Cervical Artificial Disc) [23] SECURE®-C (Globus Medical, Audubon, PA, USA) UHMWPE on CoCr N/A FE: ± 7.0°
LB: ± 7.0°
AR: ± 1.5° (Test I) and ± 6.0° (Test II)
AL: 150 N
F: 2.0 Hz
As published:
Test I: 2.57 ± 1.21 mg/MC
Test II: 0.89 ± 0.3 mg/MC
Converted (density = 0.94 mg/mm3):
Test I: 2.73 ± 1.29 mm3/MC
Test I: 0.95 ± 0.32 mm3/MC
N/A
SSED for PMA Number: P100012 (PCM® Cervical Disc) [22] PCM® (NuVasive, San Diego, CA, USA) UHMWPE on CoCr ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
6.33 mg/MC
Converted (density = 0.94 mg/mm3): 6.73 mm3/MC
Average size: 0.43 ± 0.04 µm and 0.53 ± 0.15 µm for two samples, respectively
Grupp et al., 2015 [28] activ® C (Aesculap AG, Tuttlingen, Germany) UHMWPE on CoCr ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.0 ± 0.1 mg/MC
Converted (1.29 mg/mm3):
0.78 ± 0.08 mm3/MC
99% of particles were less than 1 µm
PEEK on PEEK As published:
1.4 ± 0.4 mg/MC
Converted (1.29 mg/mm3):
1.1 ± 0.3 mm3/MC
99% of particles were less than 1 µm
CFR-PEEK on CFR-PEEK As published:
0.02 ± 0.02 mg/MC
Converted (1.29 mg/mm3):
0.02 ± 0.02 mm3/MC
99% of particles were less than 1 µm
PEK on PEK As published:
0.8 ± 0.1 mg/MC
Converted (1.29 mg/mm3):
0.6 ± 0.1 mm3/MC
90% of particles were less than 1 µm
Brown et al., 2010 [11] NuNec (Pioneer Surgical, Marquette, MI, USA) PEEK on PEEK ASTM F2423 FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 2.0 Hz
As published:
0.26 ± 0.01 mg/MC
Converted (1.29 mg/mm3):
0.20 ± 0.01 mm3/MC
N/A
ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 100 N
F: 2.0 Hz
As published:
0.32 ± 0.02 mg/MC
Converted (1.29 mg/mm3):
0.25 ± 0.02 mm3/MC
N/A
Xin et al., 2013 [51] NuNec (Pioneer Surgical) PEEK on PEEK ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 50-150 N
F: 1.0 Hz
As published:
2.5 ± 0.1 mg/MC
As published:
1.9 ± 0.1 mm3/MC
N/A
Design History File Report for the Kineflex|C Kineflex|C (SpinalMotion, Mountain View, CA) CoCr on CoCr ISO 18192-1 FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 100 N
F: 5.0 Hz
As published:
3.2 mg/ MC
Converted (Density = 8.5 mg/mm3): 0.4 mm3/MC
Average size: 0.4–0.55 µm
Kurtz et al., 2012 [35] Prestige ST (Medtronic Sofamor Danek) Stainless steel on stainless steel ASTM F2423 FE: ± 9.7° (Phase II)
LB: ± 4.7° (Phase I)
AR: ± 3.8° (Phase I)
AL: 49 N (Phase I)/149 N (Phase II)
F: 1.0 Hz
As published:
Phase I: 0.72-0.76 mm3/MC
Phase II: 0.004-0.059 mm3/MC
Average size: 0.3–0.6 µm
SSED for PMA Number: P090029 (Prestige® LP Cervical Disc) [25] Prestige® LP (Medtronic Sofamor Danek) Titanium carbide on Titanium carbide ASTM F2423* FE: ± 7.5° (Phase II)
LB: ± 6.0° (Phase I)
AR: ± 6.0° (Phase I)
AL: 100 N (Phase I/II)
F: 1.0 Hz
As published:
Phase I then Phase II: 0.35 ± 0.3 mm3/MC*
Average size less than 0.2 µm

* Four wear tests were described in the SSED for the Prestige® LP Cervical Disc; the maximum standard wear rate has been reported; ASTM = American Society for Testing and Materials; ISO = International Standards Organization; FE = flexion/extension; LB = lateral bending; AR = axial rotation; AL = axial load; F = frequency; SSED = Summary of Safety and Effectiveness Data; N/A = not available.