Table 1.

Evaluation of NINA-SUD devices

Patient ID HIV Status Viral Load (Copies/mL) NINA-SUDa Thermal cyclerb
1 UW5 Positive 230700 3/3 3/3
2 UW17 Positive 161500 3/3 3/3
3 UW29 Positive 128800 3/3 3/3
4 UW15 Positive 57120 3/3 3/3
5 UW24 Positive 50370 3/3 3/3
6 UW13 Positive 39300 3/3 3/3
7 UW16 Positive 28980 3/3 2/3
8 UW2 Positive 22870 3/3 2/3
9 UW31 Positive 11660 3/3 2/3
10 UW7 Positive 5958 2/3 2/3
11 UW25 Positive 4798 3/3 2/3
12 UW18 Positive 3323 2/3 3/3
13 UW10 Positive 3115 3/3 0/3
14 UW12 Positive 1839 0/3 0/3
15 UW23 Positive 1837 0/3 0/3
16 BRH624078 Negative N/A 0/3 0/3
17 BRH624079 Negative N/A 0/3 0/3
18 BRH624080 Negative N/A 0/3 0/3
19 BRH624081 Negative N/A 0/3 0/3
20 BRH624082 Negative N/A 0/3 0/3
a

Number positive from 3 separate NINA-SUD devices

b