Published in final edited form as: J Am Soc Echocardiogr. 2016 Jun 11;29(9):861–870. doi: 10.1016/j.echo.2016.05.002

Table 6.

Diagnostic Performance of Echo RV Linear Dimensions for Right Ventricular Chamber Dilation on CMR*

A. ASE Guideline Recommended Linear Cutoffs
Cutoff
(mm)
Sensitivity Specificity Accuracy PPV NPV
Proximal RVOT width 35 81% 75% 75% 25% 97%
Basal RV width 41 81% 84% 84% 36% 98%
Mid RV width 35 46% 92% 88% 39% 94%
RV length 83 77% 67% 68% 20% 96%
Distal RVOT width 27 68% 80% 79% 27% 96%
B. Cohort-derived Linear Cutoffs**
AUC (95% CI) Cutoff
(mm/m2)
Sensitivity Specificity Accuracy PPV NPV
Proximal RVOT width 0.86 (0.79–0.93); p<0.001 18.7 77% 82% 82% 31% 97%
Basal RV width 0.87 (0.78–0.95); p<0.001 22.4 77% 89% 87% 42% 97%
Mid RV width 0.82 (0.73–0.92); p<0.001 16.7 77% 84% 83% 33% 97%
RV length 0.78 (0.69–0.87); p<0.001 45.8 54% 82% 79% 24% 94%
Distal RVOT width 0.77 (0.66–0.88); p<0.001 14.2 72% 82% 81% 29% 97%
*

RV dilation assessed at end-diastole on both echo and CMR

**

optimized for sensitivity using a minimum specificity cutoff of 80%