Table 2.

Microbial identification by LCM/MALDI-TOF versus classical ID/VITEK2 from OXOID SIGNAL BLOOD bottles

Microorganism Number of isolates analyzed Correct id VITEK2 Correct id MALDI-TOFa No id MALDI-TOFb Id genusc Misidentified
Gram positive 16 16 12 2 1 1
 Enterococcus faecalis 4 4 3 1
 Enterococcus faecium 2 2 2
 Staphylococcus epidermidis 3 3 3
 Staphylococcus aureus 3 3 3
 Streptococcus pyogenes 1 1 1
 Streptococcus mitis 1 1 1
 Streptococcus mutans 1 1 1d
 Streptococcus parasanguinis 1 1 1
Gram negative 25 25 25
 Escherichia coli 7 7 7
 Klebsiella pneumoniae 9 9 9
 Achromobacter xylosoxidans 1 1 1
 Proteus mirabilis 1 1 1
 Pseudomonas aeruginosa 4 4 4
 Acinetobacter lwoffii 1 1 1
 Klebsiella oxytoca 1 1 1
 Enterobacter cloacae 1 1 1
Yeasts 1 1 1
 Candida tropicalis 1 1 1
Total 42

aScore values higher than 2.0 for Bruker Biotyper software

bScore values lower than 1.5 for Bruker Biotyper software

cScore values between 2.0 and 1.7 for Bruker Biotyper software

d S. mutans was misidentified as L. plantarum