Skip to main content
. Author manuscript; available in PMC: 2016 Oct 12.
Published in final edited form as: Transfusion. 2013 Feb 27;53(10 Pt 2):2384–2398. doi: 10.1111/trf.12117

TABLE 3.

Proportion of ID-NAT WP samples detectable by HIV p24 Ag assay and on retest of several ID-NAT and MP-NAT options*

Pool size Number of 73 WP samples reactive (%) in replicate assays
HIV p24 Ag (single) Ultrio (12 replicates) Ultrio Plus (12 replicates) TaqScreen (six replicates)
IDT 33.0 (45.2) 71.4 (97.8) 71.4 (97.8) 70.8 (97.2)
MP4 67.9 (93.0) 67.4 (92.3)
MP6 64.1 (88.1)
MP8 64.7 (88.6) 65.5 (89.7)
MP16 60.2 (82.4) 62.2 (85.3)
*

Data are reported as number (%).

Number of replicates reactive divided by 12 for Ultrio and Ultrio Plus and divided by six for TaqScreen.