. 2016 Oct 14;6:35419. doi: 10.1038/srep35419

Table 1. Performance of CV, DPV and OECT in the detection of AA, DA and UA.

  Three electrode cell
OECT
CV
DPV
Epaa (V vs SCE) Sensitivityb(LOD)c Sensitivityb(LOD)c Sensitivitye Sensitivityd(LOD)c Sensitivitye
AA 0.29 0.063 ± 0.005(1 10−4) 0.00967 ± 0.0003(2 10−5) 0.193 ± 0.003 0.102 ± 0.004(7 10−5) 0.85 ± 0.03
DA 0.35 0.109 ± 0.009(3 10−4) 0.448 ± 0.006(2 10−6) 9.0 ± 0.1 1.09 ± 0.04(6 10−6) 9.1 ± 0.3
UA 0.41 0.090 ± 0.002(2 10−4) 0.065 ± 0.001(6 10−6) 1.30 ± 0.02 0.40 ± 0.01(2 10−5) 3.33 ± 0.08

aanodic peak potentials evaluated by CV for 1 mM solutions.

bexpressed in A M−1 cm−2.

cexpressed in M.

dexpressed in S M−1.

enormalized sensitivity expressed in S M−1 cm−2. The DPV and OECT sensitivities are normalized to pulse amplitude and gate area, respectively.