Table 4.

Per cent-expansion area and delivery balloon volume

SAPIEN 3 (n = 39)
SAPIEN XT (n = 55)
Under-filling (n = 6) Nominal (n = 26) Over-filling (n = 7) Under-filling (n = 12) Nominal (n = 33) Over-filling (n = 10)
Overall levels 96.0 ± 3.2 100.1 ± 5.0 107.8 ± 3.4* 91.0 ± 4.7 97.5 ± 4.1** 97.4 ± 7.2
1. THV inflow 99.0 ± 5.2 104.8 ± 5.7 112.0 ± 1.5 87.7 ± 5.7 94.3 ± 4.2 96.5 ± 8.5
2. Native annulus 95.1 ± 7.8 98.4 ± 5.5 107.0 ± 3.3 89.8 ± 5.2 95.9 ± 4.2 97.8 ± 6.1
3. Mid-THV 90.9 ± 3.5 95.8 ± 4.8*** 103.5 ± 4.8 90.2 ± 5.1 96.9 ± 5.0**** 94.8 ± 7.7
4. THV outflow 99.1 ± 3.3 101.5 ± 6.3 108.6 ± 4.7 96.4 ± 8.0 102.9 ± 6.6 100.4 ± 8.8

Values are mean ± SD.

*P < 0.001 and **P = 0.029 vs. the SAPIEN 3 nominal, ***P < 0.001 vs. the SAPIEN 3 THV-inflow, ****P = 0.0041 vs. the SAPIEN XT THV-inflow.