Table 4. Slope, standard error (SE), intercept, Pearson correlation coefficient (R2) values, limits of detection (LOD) and quantitation (LOQ), and working range of myoglobin calibration curve.

RANGE Myoglobin response (0–10 mg/mL) Myoglobin response (0–75 mg/mL)
MATRIX 50% Solution A Jersey sample Holstein sample 50% Solution A Jersey sample Holstein sample
SLOPE 58336520 50745893 56156165 92981539 73818123 98846238
SE 19010009 15462981 20733548 1650282 786991 1416302
INTERCEPT 15644753 17941811 21909458 1164209 2578088 -1104311
R2 0.9920 0.9882 0.9827 0.9967 0.9988 0.9978
LOD (mg/mL) 0.98 0.91 1.11 0.05 0.03 0.04
LOQ (mg/mL) 3.26 3.05 3.69 0.18 0.11 0.14
Working range 0.9–10 mg/mL 0.8–10 mg/mL 1.1–10 mg/mL 0.05–1 mg/mL 0.03–1 mg/mL 0.04–1 mg/mL