Table 2. Intra-day and inter-day precision and accuracy of the proposed methods.

In Human Plasma matrix LQC (20 ngmL-1) MQC (150 ngmL-1) HQC (300 ngmL-1)
Intra-day assay* Inter-day assay** Intra-day assay Inter-day assay Intra-day assay Inter-day assay
Mean 19.70 19.89 147.95 147.72 299.21 299.48
Standard deviation (SD) 0.31 0.50 1.67 1.61 3.16 3.66
Precision (%RSD) 1.58 2.54 1.13 1.09 1.06 1.22
Accuracy (%RE) -1.48 -0.55 -1.37 -1.52 -0.26 -0.17
In RLMs matrix LQC (20 ngmL-1) MQC (150 ngmL-1) HQC (300 ngmL-1)
Intra-day assay* Inter-day assay** Intra-day assay Inter-day assay Intra-day assay Inter-day assay
Mean 19.67 19.97 147.75 147.88 299.1 299.07
Standard deviation (SD) 0.29 0.46 1.59 1.86 3.08 2.90
Precision (%RSD) 1.49 2.31 1.07 1.26 1.03 0.97
Accuracy (%RE) -1.65 -0.16 -1.05 -1.42 -0.3 -0.31

* Average of twelve determinations of day 1.

** Average of six determinations in three consecutive days.