Table 4. PNT Stability data and dilution integrity in plasma matrix under different conditions.
Nominal Concentration(ng.mL−1) | Mean (ngmL−1) | Recovery % | Precision (CV %) | Accuracy (RE %) | |
---|---|---|---|---|---|
Room Temp. for 8h | |||||
20 | 20.28 | 101.40 | 2.12 | 1.41 | |
150 | 147.77 | 98.51 | 1.37 | -1.49 | |
300 | 298.54 | 99.51 | 0.67 | -0.49 | |
Three freeze-thaw cycles | |||||
20 | 19.81 | 99.05 | 2.38 | -0.94 | |
150 | 147.86 | 98.57 | 1.09 | -1.43 | |
300 | 299.61 | 99.87 | 1.22 | -0.13 | |
Stored at 4 o C for 24 h | |||||
20 | 20.29 | 101.45 | 2.17 | 1.46 | |
150 | 147.87 | 98.58 | 1.57 | -1.42 | |
300 | 298.83 | 99.61 | 1.01 | -0.39 | |
Stored at -20 o C for 30 days | |||||
20 | 19.62 | 98.10 | 1.65 | -1.90 | |
150 | 147.76 | 98.51 | 1.21 | -1.49 | |
300 | 298.93 | 99.64 | 0.96 | -0.36 | |
Dilution integrity | |||||
180 | 176.49 | 98.05 | 1.63 | -1.95 | |
360 | 358.28 | 99.52 | 1.77 | -0.48 |