Skip to main content
. 2016 Oct 10;2016:3584137. doi: 10.1155/2016/3584137

Table 2.

The Bland-Altman 95% limits agreement and intraclass correlation coefficient for spherical equivalent.

Parameter Mean difference (standard deviation) Bland-Altman
95% limits of agreement
p value ICC (95% confidence interval)
Autorefraction versus retinoscopy 0.26 (1.67) −3.01 to 3.53 0.348 0.902 (0.809–0.950)
Subjective versus retinoscopy 0.12 (0.51) −0.88 to 1.12 0.157 0.989 (0.978–0.994)
Subjective versus autorefraction 0.37 (1.64) −2.84 to 3.58 0.167 0.895 (0.796–0.947)

ICC = intraclass correlation coefficient.