Skip to main content
. 2016 Oct 26;188(11):640. doi: 10.1007/s10661-016-5647-y

Table 4.

Use of FAAS and ICP-MS techniques for determination of thallium in water and biological material samples (Li et al. 2009)

Sample ICP-MS technique FAAS technique Relative standard deviation [%] Recovery [%]
Tap water 0.260 μg L−1 0.240 μg L−1 3.2 88–102
River water 1.220 μg L−1 1.340 μg L−1 3.4 91–106
Human hair 0.384 mg kg−1 0.395 mg kg−1 3.0 90–98
Human nail 0.675 mg kg−1 0.668 mg kg−1 3.2 88–97