Abstract
Objectives
Fetal cardiovascular magnetic resonance imaging (MRI) offers a potential alternative to echocardiography, although in practice, its use has been limited. We sought to explore the need for additional imaging in a tertiary fetal cardiology unit and the usefulness of standard MRI sequences.
Methods
Cases where the diagnosis was not fully resolved using echocardiography were referred for MRI. Following a three‐plane localiser, fetal movement was assessed with a balanced steady‐state free precession (bSSFP) cine. Single‐shot fast spin echo and bSSFP sequences were used for diagnostic imaging.
Results
Twenty‐two fetal cardiac MRIs were performed over 12 months, at mean gestation of 32 weeks (26–38 weeks). The majority of referrals were for suspected vascular abnormalities (17/22), particularly involving the aortic arch (n = 10) and pulmonary vessels (n = 4). Single‐shot fast spin echo sequences produced ‘black‐blood’ images, useful for examining the extracardiac vasculature in these cases. BSSFP sequences were more useful for intracardiac structures. Real‐time SSFP allowed for dynamic assessment of structures such as cardiac masses, with enhancement patterns also allowing for tissue characterisation in these cases.
Conclusions
Fetal vascular abnormalities such as coarctation can be difficult to diagnose by using ultrasound. Fetal MRI may have an adjunctive role in the evaluation of the extracardiac vascular anatomy and tissue characterisation. © 2016 The Authors. Prenatal Diagnosis published by John Wiley & Sons, Ltd.
Short abstract
What's Already Known About This Topic?
Fetal cardiac MRI offers the potential to be a safe adjunct to echocardiography; however, there is no consensus on routine fetal indications or technical protocols for cardiovascular MRI in clinical practice. Previous studies have focused mainly on its utility for intracardiac imaging.
What Does This Study Add?
Abnormalities involving the extracardiac vasculature were the most common group of referrals for additional imaging in clinical practice. Fast spin echo MRI sequences offer the potential for better visualisation of these structures than ultrasound alone. Additional MRI benefits such as tissue characterisation can add further value in selected cases.
Introduction
Conventional fetal echocardiography is the mainstay for prenatal diagnosis of congenital cardiac lesions, based on its ease of use, availability and high diagnostic accuracy.1, 2, 3 However, individual fetal and maternal factors can have a deleterious effect on the quality of ultrasound imaging,4, 5 and there remain inherent difficulties in diagnosing specific forms of congenital heart disease in the fetus.3, 6, 7, 8 Despite this, due to technical and safety considerations, alternatives to ultrasonographic techniques have traditionally been extremely limited.9
Magnetic resonance imaging (MRI) has been an established adjunct for postnatal assessment of the cardiovascular system since the 1980s,10 with routine studies able to deliver three‐dimensional angiography, detailed real‐time imaging, cardiac volumes and vascular flow measurements.11 Cardiovascular MR imaging in the fetus, however, presents several challenges not present after birth. The fetal heart is extremely small, with a total length of between 2 and 4 cm depending on gestational age,12 and beats with high heart rates of up to 160 beats per minute.13 Electrocardiographic ‘gating’, a means of improving spatiotemporal resolution, is not yet practicable in the fetus, although several non‐invasive methods are under investigation.14, 15 Unpredictable three‐dimensional fetal motion, maternal bulk motion and fetal and maternal breathing movements present additional difficulties,16 with no additional benefit from attempted fetal sedation.5
The first report of attempted fetal cardiac MRI was in 2005, using real‐time sequences with an attempt to quantify ventricular volumes.17 There followed a number of small case series of fetuses with normal hearts18, 19 and congenital heart disease.20 Whilst several subsequent reviews have attempted to develop defined protocols,21, 22 few have focused on the clinical role and utility of fetal cardiac MRI,5 and there is currently no consensus on routine fetal indications or technical protocols for cardiovascular MRI in clinical practice.
The increasing use of MRI for prenatal evaluation of the fetal brain23 and other organs24 offers the potential for more widespread access to the expertise and infrastructure required to deliver fetal MRI programmes. We sought to explore both the need for adjunctive fetal cardiac imaging in a tertiary fetal cardiology unit and the usefulness of standard MRI sequences in delivering additional diagnostic information.
Aims
Over a one‐year period, fetal MRI was requested on clinical grounds by the fetal cardiology service, reserved for any difficult fetal cases where the diagnosis was in doubt or could benefit from additional imaging. We present an analysis of the pattern of referrals we subsequently observed and the diagnostic utility of standard MRI sequences in this context.
Methods
Referrals were based on the judgement of the attending fetal cardiologists (JS, OM, VZ and GS). All examinations took place between June 2014 and May 2015. Pre‐existing ethical approval was in place for a long‐standing project examining brain growth in fetuses with CHD, which also permitted the use of cardiac sequences (REC: 07/H0707/105). All participants provided written consent and received concomitant fetal brain imaging. The timing of the MRI was determined by the clinical teams, taking into account the gestational age at diagnosis, the potential impact of new information on fetal counselling and management, and the timeframe for optimal imaging quality (generally later gestation, when the fetus is larger with fewer major translational movements). A paediatric cardiologist with expertise in cardiac MR was present at every study (RR or KP). Each mother's tympanic temperature was taken before and after the MRI or if they became uncomfortable during the study. Conservative limits of a rise in temperature of ≥1 °C and/or an absolute estimated temperature in the fetus of greater than 38 °C (i.e. maternal temperature +0.5 °C)17 were set; breach of either would lead to the study being discontinued.
A large field of view (covering the whole fetus) bSSFP cine (real‐time) scan with low spatial resolution but higher temporal resolution (303 ms) was used following the initial localiser sequence to demonstrate the degree of gross fetal movement.25 Diagnostic sequences were generally limited to half‐Fourier acquisition single‐shot fast spin‐echo (‘SSFSE’) and balanced steady‐state free precession (‘bSSFP’) gradient echo sequences, as these sequences were felt to offer reasonable signal contrast and spatiotemporal resolution within established limits of radiofrequency power deposition26 and acoustic noise27. These were either multi‐slice imaging to provide whole heart coverage using parallel overlapping slices or single‐slice imaging in targeted scan planes where dynamic information was of interest (real‐time; Table 1). If needed, further shortened localiser sequences were also used to account for fetal movement. Sequences were repeated as needed to a maximum study time of 60 min.
Table 1.
Sequence | Multi‐slice SSFSE | Multi‐slice bSSFP | Single‐slice bSSFP |
---|---|---|---|
TR/TE (ms) | 15 000/100 | 4.3/2.1 | 3.9/1.9 |
Flip angle (degrees) | 90 | 90 | 60 |
Field of view (mm) | 350 × 350 | 350 × 350 | 400 × 300 |
In plane resolution (mm) | 1.4 × 1.4 | 1.5 × 2.3 | 1.8 × 2.3 |
Slice thickness (mm) | 2.5 | 5 | 5 |
Order | 16 interleaved/TR | Linear slices | Dynamic single slice |
SENSE factor | 2 | 1 | 2 |
Partial‐Fourier factor | 5/8 | 1 | 5/8 |
Single‐slice duration (ms) | 468 | 647 | 159 |
SSFSE, single shot fast spin echo; bSSFP, balanced steady‐state free precession; TR, repetition time; TE, echo time; SENSE, sensitivity encoding.
Results
Twenty‐two fetal cardiac MRI examinations were performed over the referral period. The average gestational age at diagnosis 23+4 weeks (median 23+3 weeks, range 15+4–34+0 weeks) and the average age at MRI was 31+3 weeks (median 33+2 weeks, range 22+2–38+1 weeks). Two twin pregnancies were included, each with one healthy and one affected twin.
The most common indications were for possible coarctation of the aorta in six patients, other arch abnormalities in four patients (including aortic interruption in two patients, right aortic arch and aberrant left subclavian artery) and assessment of pulmonary vessels in four patients (Table 2). In total, referrals for clarification of the extracardiac cardiovascular anatomy comprised 77% of all referrals (17/22 cases). No examinations were abandoned early for maternal claustrophobia. One patient experienced a rise in temperature to 38.3 °C towards the end of the study, which was then discontinued.
Table 2.
Indication | N | MRI successful | No cardiac data obtained | No postnatal data |
---|---|---|---|---|
Possible coarctation | 6 | 5 | 1a | — |
Aortic arch anatomy | 4 | 3 | 1a | — |
Pulmonary vasculature | 4 | 3 | 1b | — |
Other vascular | 3 | 3 | — | 1 (TOP 35w) |
Cardiac mass | 3 | 3 | — | 1 (TOP 32w) |
Cardiac diverticulum | 2 | 2 | — | — |
TOTAL (%) | 22 (100%) | 19 (86%) | 3 (14%) | 2 (9%) |
TOP, termination of pregnancy.
Excessive fetal movement.
Unable to visualise pulmonary veins due to lung compression from large hernia.
Three examinations failed to provide any useful cardiovascular information. In two cases, this was due to excessive fetal movement (at 29 and 33 weeks). In one case, assessment of the pulmonary veins in a 26‐week fetus was impossible due to severe lung compression from a large diaphragmatic hernia; however, this important prognostic information was used to help counsel the parents prior to the demise of the fetus at 29 weeks. No postnatal data were available in two cases (one intrauterine death at 32 weeks and one medical termination at 35 weeks). Autopsy was not performed in either case. In all remaining patients, the MRI diagnosis was confirmed postnatally. A comprehensive summary of all studies performed is shown in Table 3.
Table 3.
GA | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
n | W | D | Fetal diagnosis (USA) | MRI indication | Co‐morbidity | MRI findings | Additional MRI benefits | Postnatal diagnosis | Outcome | Age |
1 | 29 | 2 |
AVSD Ventricular asymmetry |
Evidence of CoA? | Cerebral ventriculomegaly | Abandoned: excessive fetal movement | Concomitant MRI brain |
Borderline left heart Coarctation of the aorta (echocardiography and surgery) |
Hybrid procedure | 5 days |
2 | 32 | 2 | Possible coarctation | Evidence of CoA? | None |
Hypoplasia of the aortic isthmus ‘Posterior shelf’ (Figure 1) Ventricular asymmetry LV < RV High risk of coarctation |
— | Coarctation of the aorta (echocardiography and surgery) | Coarctation repair | 3 days |
3 | 35 | 3 | Hypoplastic left heart syndrome | Pulmonary venous anatomy? | None | Large, tortuous, solitary pulmonary veins bilaterally (Figure 3) | — | As per prenatal imaging (postnatal MRI and surgery) | Hybrid procedure | 3 days |
4 | 33 | 3 | ?Right ventricular aneurysm Small pericardial effusion | Ventricular aneurysm/ diverticulum? | None |
Normal cardiac connections No aneurysm seen |
— | Structurally normal heart (echocardiography) | Outpatient review | — |
5 | 34 | 0 |
Hypoplastic left heart syndrome Restrictive atrial septum |
Pulmonary venous anatomy? | None | Dilated pulmonary veins with normal anatomy | No pulmonary lymphangectasia | As per prenatal imaging (echocardiography and surgery) | Hybrid procedure | 4 days |
6 | 34 | 0 |
Muscular VSD Possible coarctation |
Evidence of CoA? | None |
Mild ventricular asymmetry Normal arch and isthmus Low suspicion of coarctation |
— | Structurally normal heart (echocardiography) | Outpatient review | — |
7 | 33 | 2 | Left atrial mass | Nature of cardiac mass? | Possible tuberous sclerosis | Left atrial mass: atrial myxoma, haemangioma or rhabdomyoma (Figure 7) | Tissue characterisation | Left atrial rhabdomyoma (histology) | Surgical resection | — |
8 | 38 | 1 | RV apex tumour | Nature of cardiac mass? | None | Solitary RV rhabdomyoma (Figure 8) |
Tissue characterisation Concomitant MRI brain |
As per prenatal imaging (echocardiography) | Outpatient review | — |
9 | 33 | 5 |
Mal‐aligned VSD Interrupted aortic arch |
Aortic arch anatomy? | 22q11 microdeletion (postnatal) |
Type B interruption Subclavian in continuity with descending Ao |
— | As per prenatal imaging (echocardiography and surgery) | Surgical repair | 2 days |
10 | 33 | 6 |
Common arterial trunk ?Interruption of aortic arch |
Aortic arch anatomy? |
22q11 microdeletion Left hydronephrosis |
Common arterial trunk with arterial duct No interruption |
— | As per prenatal imaging (postnatal MRI and surgery) | Surgical repair | 9 days |
11 | 28 | 2 |
Twin 1: normal Twin 2: multiple cardiac masses |
Nature of cardiac masses? | DCDA twins |
Twin 1: normal Twin 2: three intramyocardial tumours suggestive of rhabdomyomas on T2 weighting |
Tissue characterisation MRI brain: multiple rhabdomyomas Twin 1: normal heart |
Selective feticide: no post‐mortem Maternal germ‐line mosaicism for tuberous sclerosis |
Twin 1 well, twin 2: selective feticide | 32/40 |
12 | 32 | 1 |
?Coarctation of the aorta Bilateral SVCs |
Evidence of CoA? | None |
Hypoplasia of the aortic isthmus Long gap LCCA–LSA High‐risk neonatal coarctation (Figure 4) |
— | Mild coarctation of the aorta only (Figure 5) (postnatal MRI) | No intervention to date | 29 days |
13 | 29 | 6 |
Dysplastic pulmonary valve Severe pulmonary regurgitation Maternal obesity |
Pulmonary artery anatomy? | None | Confluent right and left PAs | — | As per prenatal imaging (postnatal CT) | Outpatient review | — |
14 | 33 | 4 |
Right‐sided aortic arch Left‐sided arterial duct ALSA |
Aortic arch anatomy? | None | Inadequate views (fetal motion) | — | Right‐sided aortic arch ALSA (echocardiography) | Outpatient review | — |
15 | 33 | 5 |
Small aortic arch ?Coarctation of the aorta |
Evidence of CoA? | None |
Moderate transverse arch hypoplasia High‐risk coarctation |
— | Coarctation of the aorta (echocardiography and surgery) | Surgical repair | 6 days |
16 | 36 | 5 |
Twin 1: R arch, L duct, ALSA Twin 2: normal |
Aortic arch anatomy? | None |
Twin 1: R arch, L duct, ALSA Twin 2: normal |
No airway compression | As per prenatal imaging (postnatal MRI) | Outpatient review | — |
17 | 33 | 1 |
Cardiomegaly, RV>LV Dilated great vessels Normal aortic isthmus |
Other vascular: exclude large AVMs | None |
Large RV, PA and Ao Normal connections No AVMs |
— | As per antenatal imaging (echocardiography) | Outpatient review | — |
18 | 26 | 4 | Diaphragmatic hernia | Pulmonary venous anatomy? | Diaphragmatic hernia | Not able to visualise pulmonary veins due to lung compression | Large diaphragmatic hernia with pulmonary hypoplasia |
Normal pulmonary veins Absent diaphragm bilaterally (post‐mortem) |
Medical termination | 29/40 |
19 | 30 | 3 | Possible RV diverticulum | Ventricular aneurysm/ diverticulum? | SCD | Wide‐mouthed contractile diverticulum at base of right ventricle, 10 × 12 mm (Figure 6) | — | As per prenatal imaging (echocardiography) | Outpatient review | — |
20 | 28 | 0 | Abnormal arterial duct | Other vascular: arterial duct | None |
Leftward rotation of cardiac axis with tortuous arterial duct Normal vascular anatomy |
— | Structurally normal heart (echocardiography) | Outpatient review | |
21 | 22 | 4 |
Bilateral SVCs, apex to right ?Other vascular abnormality |
Other vascular | Jacobsen syndrome |
Rightward rotation of cardiac axis Bilateral SVCs Normal vascular anatomy |
— |
Medical termination Genetics: Jacobsen syndrome (no post‐mortem) |
Medical termination | 35/40 |
22 | 22 | 2 |
Bilateral SVCs ?Coarctation of the aorta |
Evidence of CoA? | Deficient cerebellar vermis | Dominant RV, bilateral SVCs, hypoplastic aortic arch, high probability of coarctation | Deficient cerebellar vermis diagnosed | Coarctation of the aorta (echocardiography) |
Premature Died 1 day of age (atrial arrhythmia) |
33/40 |
AVSD, atrioventricular septal defect; CoA, coarctation of the aorta; LV, left ventricle, RV, right ventricle; VSD, ventricular septal defect; LCCA, left common carotid artery; LSA, left subclavian artery; PA, pulmonary artery; ALSA, aberrant left subclavian artery; LPA, left pulmonary artery; DCSA, doubly committed sub‐arterial; AVM, arteriovenous malformation; Ao, aorta; CHAOS, congenital high airway obstruction syndrome; SCD, sickle cell disease.
MRI sequences
Single‐shot fast spin echo sequences produced T2‐weighted images with black‐blood‐like contrast with excellent contrast between the vasculature and higher signal intensity of the thymus, lung and myocardium. These sequences were the most useful for assessment of extra‐cardiac vascular anatomy as these structures are less mobile throughout the cardiac cycle (Figures 1, 2, 3, 4). They were, however, still prone to artefact from gross fetal movement. It is important to note that three‐dimensional interrogation of multi‐slice sequences was not possible, as even small fetal movements produced non‐contiguous anatomy between slices (see the Future directions section). An example of a standard postnatal ‘black‐blood’ sequence is shown in Figure 5.
Balanced SSFP images produced bright blood images with good contrast between the blood pool and surrounding tissue, particularly intracardiac structures (Figures 6, 7, 8); however, they were more susceptible to motion and/or flow artefacts. Real‐time bSSFP allowed for dynamic imaging of the beating fetal heart and was the most reliable sequence for assessment of moving structures, i.e. aneurysms and diverticulums (Figure 6). Differences in enhancement patterns between sequences allowed for tissue characterisation of intracardiac masses by using established MRI techniques.28
Discussion
We instituted an open referral process for fetal MRI in a tertiary UK fetal cardiac centre. Whilst previous explorations of the clinical utility of fetal cardiac MRI have focused predominantly on intracardiac anatomy,5 the most common indication in our series was clarification of the extracardiac vascular anatomy, in keeping with established evidence that abnormalities involving these vessels, such as the coarctation of the aorta8, 29, 30 and anomalies of the pulmonary veins,6, 7 are amongst the most challenging to define using ultrasound alone. Novel 3D and 4D echocardiographic techniques, such as spatiotemporal image correlation, have shown some potential for adjunctive imaging and can be visualised in three dimensions, but are still subject to many of the same limitations as conventional ultrasound.12 Promisingly, however, and in keeping with animal models,31 our initial data suggest a promising strength of fetal MRI in vascular imaging, particularly using SSFSE sequences. To date, the only other report examining the clinical utility of fetal cardiac MRI used bSSFP sequences only, at a similar median gestational age to our series, interpreted blindly in 40 fetuses with CHD. Whilst this showed reasonable sensitivity for detection of abnormal intracardiac anatomy, the authors concede that MRI added no new diagnostic information when compared to ultrasound. In the same study, however, 15 fetuses were classified postnatally as having an abnormal aortic arch, of which five had been missed on ultrasound, six were missed on MRI and four were detected on both.5 Unlike in our study, no SSFSE sequences were used for vascular imaging.
The single most frequent single referral indication in our series was for suspected coarctation of the aorta, a notoriously challenging condition to predict prenatally due to both the technical difficulty of imaging this region and the obligatory patency of the arterial duct in fetal life, the postnatal closure of which can only then reveal life‐threatening aortic constriction.7, 8, 30 Alongside extracardiac and genetic associations, a number of anatomical fetal risk factors for the condition have been suggested, such as the absolute and relative diameter of aortic isthmus, the presence of a posterior shelf, ventricular/great artery asymmetry and the persistence of a left‐sided superior caval vein.8 In fetal MRI, imaging quality is unaffected by artefact from the fetal spine and ribcage, offering the potential for more accurate visualisation of these subtle anatomical factors. It is well established, however, that even a detailed anatomical assessment is not sufficient to accurately predict postnatal coarctation; for example, one of our cases with clearly visualised hypoplasia of the aortic arch before birth has gone on to develop only a mild form of coarctation, which has yet to require intervention (Figures 4 and 5). Indeed, it is likely that coarctation has a particularly complex aetiology, not yet fully understood, related to intrinsic myocardial abnormalities and/or a reduction in flow through the left heart. This has only recently started to be explored by using functional ultrasound techniques, rather than relying on anatomical findings alone.32 The addition of more advanced MRI techniques in the future may help to define a more robust method of stratifying fetuses with suspected coarctation, providing a more detailed physiological assessment of affected fetuses (see the Future directions section).
In addition to vascular imaging, in three cases the combined use of SSFSE sequences provided the unique ability to characterise cardiac tumours before birth, a potential use of prenatal MRI with a high prognostic impact.33 Multi‐slice and real‐time bSSFP sequences were most useful for intracardiac imaging, particularly for demonstrating the anatomy of gross lesions such as cardiac masses and diverticulums.
Finally, it should also be noted that several of the MRI studies reported in our series concomitantly provided important complementary imaging that could directly or indirectly impact the cardiac prognosis, for example describing the devastating extent of pulmonary compression from a very large diaphragmatic hernia in a fetus that did not survive to term. Fetal MRI can also be used to assess lung parenchymal lesions and airway compromise in CHD34 and has been used to demonstrate the presence of pulmonary lymphangectasia in a fetus with hypoplastic left heart syndrome, an important prognostic finding.35 MRI of the fetal brain may provide crucial information to clinicians and parents regarding medium‐term and long‐term prognoses, which may have a profound effect on both prenatal and postnatal management independent of the underlying cardiac diagnosis.23, 36, 37
Future directions
As we observed, even small fetal movements can disrupt the spatial relationship of multi‐slice sequences, making post‐hoc analysis extremely challenging. The application of novel motion‐corrected slice‐volume registration algorithms to multiple MRI image stacks has been used to generate accurate 3D volumes of the fetal thorax.38 These exciting techniques have the potential to further enhance the capability of fetal MRI for vascular imaging, both by improving image quality and reducing the gestation age at which useful MR data can be obtained, and are currently under investigation by our unit.39
Other advanced MRI techniques can offer novel means of evaluation of physiological parameters of the fetal heart and circulation. The application of retrospective ‘metric‐optimised’ gating to phase contrast sequences can be used to quantify vascular flows and is currently being explored in a wide variety of settings.40, 41, 42, 43 Related technology allows for the retrospective gating of single slice real‐time sequences to dramatically improve the spatiotemporal resolution for cardiac imaging.44 The potential extension of these real‐time sequences to whole‐heart coverage allows for the possibility of performing comprehensive volumetric measurements throughout the cardiac cycle, providing estimates of chamber volumes and ventricular function.
These innovations will not only help to bring prenatal cardiovascular MRI closer to its postnatal counterpart but could also offer the potential to enhance our fundamental understanding of the nature and aetiology of congenital heart defects – including challenging prenatal diagnoses such as coarctation of the aorta – and their interactions with other organ systems such as the placenta45 and the developing fetal brain.37 The incorporation of all of the above techniques is the subject of ongoing research within our departments.
Conclusion
Abnormalities of the fetal vasculature such as coarctation of the aorta can be challenging to diagnose with ultrasound, even in experienced hands. Standard sequence MRI may have an adjunctive role in the assessment of the extracardiac vascular anatomy and tissue characterisation. Developments such as motion‐corrected slice‐volume registration and metric‐optimised gated phase contrast flow quantification and real‐time techniques, alongside increasingly accessible infrastructure, may further develop the role of MRI as an adjunct to echocardiography in the investigation and diagnosis of congenital heart disease in the fetus.
Limitations
We have described 22 cases that were selected for additional imaging from a single institution. Due to the increased likelihood of image corruption from uncontrolled fetal movement, we tended to perform additional imaging in the third trimester. Whilst our median gestational age of 33 weeks is comparable to other published series, this does represent a current limitation of the modality. As imaging was performed for clinical evaluation, the acquisition and interpretation of MRI images was not blinded to previous assessments. In two fetuses that did not survive to term, no autopsy information was available due to lack of parental consent.
What's Already Known About This Topic?
Fetal cardiac MRI offers the potential to be a safe adjunct to echocardiography; however, there is no consensus on routine fetal indications or technical protocols for cardiovascular MRI in clinical practice. Previous studies have focused mainly on its utility for intracardiac imaging.
What Does This Study Add?
Abnormalities involving the extracardiac vasculature were the most common group of referrals for additional imaging in clinical practice. Fast spin echo MRI sequences offer the potential for better visualisation of these structures than ultrasound alone. Additional MRI benefits such as tissue characterisation can add further value in selected cases.
Lloyd, D. F. A. , van Amerom, J. F. P. , Pushparajah, K. , Simpson, J. M. , Zidere, V. , Miller, O. , Sharland, G. , Allsop, J. , Fox, M. , Lohezic, M. , Murgasova, M. , Malamateniou, C. , Hajnal, J. V. , Rutherford, M. , and Razavi, R. (2016) An exploration of the potential utility of fetal cardiovascular MRI as an adjunct to fetal echocardiography. Prenat Diagn, 36: 916–925. doi: 10.1002/pd.4912.
Funding sources: The authors acknowledge financial support from the Department of Health via the National Institute for Health Research comprehensive Biomedical Research Centre award to Guy's and St Thomas' NHS Foundation Trust in partnership with King's College London and King's College Hospital NHS Foundation Trust, and from the iFind Project (Wellcome Trust IEH Award 102431).
Conflict of interest: None
References
- 1. Allan LD, Sharland GK, Milburn A, et al Prospective diagnosis of 1,006 consecutive cases of congenital heart disease in the fetus. J Am Coll Cardiol 1994;23:1452–8. [DOI] [PubMed] [Google Scholar]
- 2. Allan L. Technique of fetal echocardiography. Pediatr Cardiol 2004;25:223–33. [DOI] [PubMed] [Google Scholar]
- 3. Hunter LE, Simpson JM. Prenatal screening for structural congenital heart disease. Nat Rev Cardiol 2014;11:323–34. [DOI] [PubMed] [Google Scholar]
- 4. Hendler I, Blackwell SC, Bujold E, et al The impact of maternal obesity on midtrimester sonographic visualization of fetal cardiac and craniospinal structures. Int J Obes Relat Metab Disord 2004;28:1607–11. [DOI] [PubMed] [Google Scholar]
- 5. Votino C, Jani J, Damry N, et al Magnetic resonance imaging in the normal fetal heart and in congenital heart disease. Ultrasound in Obstetrics & Gynecology : The Official Journal of the International Society of Ultrasound in Obstetrics and Gynecology 2012;39:322–9. [DOI] [PubMed] [Google Scholar]
- 6. Seale AN, Carvalho JS, Gardiner HM, et al Total anomalous pulmonary venous connection: impact of prenatal diagnosis. Ultrasound Obstet Gynecol 2012;40:310–8. [DOI] [PubMed] [Google Scholar]
- 7. Simpson JM. Impact of fetal echocardiography. Ann Pediatr Cardiol 2009;2:41–50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Jowett V, Aparicio P, Santhakumaran S, et al Sonographic predictors of surgery in fetal coarctation of the aorta. Ultrasound Obstet Gynecol 2012;40:47–54. [DOI] [PubMed] [Google Scholar]
- 9. Deng, J & Rodeck, CH . Current applications of fetal cardiac imaging technology. Current Opinion in Obstetrics and Gynecology, 2006. [DOI] [PubMed]
- 10. Goldman MR, Pohost GM, Ingwall JS, et al Nuclear magnetic resonance imaging: potential cardiac applications. Am J Cardiol 1980;46:1278–83. [DOI] [PubMed] [Google Scholar]
- 11. Pennell DJ. Cardiovascular magnetic resonance. Circulation 2010;121:692–705. [DOI] [PubMed] [Google Scholar]
- 12. Sharland GK, Allan LD. Normal fetal cardiac measurements derived by cross‐sectional echocardiography. Ultrasound Obstet Gynecol 1992;2:175–81. [DOI] [PubMed] [Google Scholar]
- 13. Pildner von Steinburg S, Boulesteix AL, Lederer C, et al What is the “normal” fetal heart rate? Peer J 2013;1:e82. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Jansz MS, Seed M, van Amerom JFP, et al Metric optimized gating for fetal cardiac MRI. Magn Reson Med 2010;64:1304–14. [DOI] [PubMed] [Google Scholar]
- 15. Yamamura J, Kopp I, Frisch M, et al Cardiac MRI of the fetal heart using a novel triggering method: initial results in an animal model. J Magn Reson Imaging 2012;35:1071–6. [DOI] [PubMed] [Google Scholar]
- 16. Malamateniou C, Malik SJ, Counsell SJ, et al Motion‐compensation techniques in neonatal and fetal MR imaging. AJNR Am J Neuroradiol 2013;34:1124–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Fogel MA, Wilson RD, Flake A, et al Preliminary investigations into a new method of functional assessment of the fetal heart using a novel application of ‘real‐time’ cardiac magnetic resonance imaging. Fetal Diagn Ther 2005;20:475–80. [DOI] [PubMed] [Google Scholar]
- 18. Gorincour G, Bourliere‐Najean B, Bonello B, et al Feasibility of fetal cardiac magnetic resonance imaging: preliminary experience. Ultrasound Obstet Gynecol 2007;29:105–8. [DOI] [PubMed] [Google Scholar]
- 19. Manganaro L, Savelli S, Di Maurizio M, et al Potential role of fetal cardiac evaluation with magnetic resonance imaging: preliminary experience. Prenat Diagn 2008;28:148–56. [DOI] [PubMed] [Google Scholar]
- 20. Manganaro L, Savelli S, Di Maurizio M, et al Assessment of congenital heart disease (CHD): is there a role for fetal magnetic resonance imaging (MRI)? Eur J Radiol 2009;72:172–80. [DOI] [PubMed] [Google Scholar]
- 21. Saleem SN. Feasibility of MRI of the fetal heart with balanced steady‐state free precession sequence along fetal body and cardiac planes. AJR Am J Roentgenol 2008;191:1208–15. [DOI] [PubMed] [Google Scholar]
- 22. Manganaro L, Di Maurizio M & Savelli S. Feasibility, Technique and Potential Role of Fetal Cardiovascular MRI: Evaluation of Normal Anatomical Structures and Assessment of Congenital Heart Disease Four Dimensional Fetal Echocardiography 2010;178–196. 2010. [Google Scholar]
- 23. Rutherford M, Jiang S, Allsop J, et al MR imaging methods for assessing fetal brain development. Dev Neurobiol 2008;68:700–11. [DOI] [PubMed] [Google Scholar]
- 24. Reddy UM, Abuhamad AZ, Levine D, et al Fetal imaging: executive summary of a Joint Eunice Kennedy Shriver National Institute of Child Health and Human Development, Society for Maternal‐Fetal Medicine, American Institute of Ultrasound in Medicine, American College of Obstetricians and Gynecologists, American College of Radiology, Society for Pediatric Radiology, and Society of Radiologists in Ultrasound Fetal Imaging Workshop. Am J Obstet Gynecol 2014;210:387–97. [DOI] [PubMed] [Google Scholar]
- 25. Hayat TT, Nihat A, Martinez‐Biarge M, et al Optimization and initial experience of a multisection balanced steady‐state free precession cine sequence for the assessment of fetal behavior in utero. AJNR Am J Neuroradiol 2011;32:331–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Hand JW, Li Y, Hajnal JV. Numerical study of RF exposure and the resulting temperature rise in the foetus during a magnetic resonance procedure. Phys Med Biol 2010;55:913–30. [DOI] [PubMed] [Google Scholar]
- 27. De Wilde JP, Rivers AW, Price DL. A review of the current use of magnetic resonance imaging in pregnancy and safety implications for the fetus. Prog Biophys Mol Biol 2005;87:335–53. [DOI] [PubMed] [Google Scholar]
- 28. O'Donnell DH, Abbara S, Chaithiraphan V, et al Cardiac tumors: optimal cardiac MR sequences and spectrum of imaging appearances. Am J Roentgenol 2009;193:377–87. [DOI] [PubMed] [Google Scholar]
- 29. Franklin O, Burch M, Manning N, et al Prenatal diagnosis of coarctation of the aorta improves survival and reduces morbidity. Heart 2002;87:67–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Head CE, Jowett VC, Sharland GK, et al Timing of presentation and postnatal outcome of infants suspected of having coarctation of the aorta during fetal life. Heart 2005;91:1070–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Yamamura J, Schnackenburg B, Kooijmann H, et al Magnetic resonance angiography of fetal vessels: feasibility study in the sheep fetus. Jpn J Radiol 2010;28:720–6. [DOI] [PubMed] [Google Scholar]
- 32. Miranda, JO , Hunter, L , Tibby, S , et al. Myocardial deformation in fetuses with coarctation of the aorta: a case–control study. Ultrasound in Obstetrics & Gynecology, 2016: (online first). [DOI] [PubMed]
- 33. Kivelitz DE, Muhler M, Rake A, et al MRI of cardiac rhabdomyoma in the fetus. Eur Radiol 2004;14:1513–6. [DOI] [PubMed] [Google Scholar]
- 34. Sun HY, Boe J, Rubesova E, et al Fetal MRI correlates with postnatal CT angiogram assessment of pulmonary anatomy in Tetralogy of Fallot with absent pulmonary valve. Congenit Heart Dis 2014;9:E105–9. [DOI] [PubMed] [Google Scholar]
- 35. Seed M, Bradley T, Bourgeois J, et al Antenatal MR imaging of pulmonary lymphangiectasia secondary to hypoplastic left heart syndrome. Pediatr Radiol 2009;39:747–9. [DOI] [PubMed] [Google Scholar]
- 36. Limperopoulos C, Tworetzky W, McElhinney DB, et al Brain volume and metabolism in fetuses with congenital heart disease: evaluation with quantitative magnetic resonance imaging and spectroscopy. Circulation 2010;121:26–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Brossard‐Racine M, du Plessis A, Vezina G, et al Brain injury in neonates with complex congenital heart disease: what is the predictive value of MRI in the fetal period? AJNR Am J Neuroradiol 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Kainz B, Malamateniou C, Murgasova M, et al Motion corrected 3D reconstruction of the fetal thorax from prenatal MRI. Med Image Comput Comput Assist Interv 2014;17:284–91. [DOI] [PubMed] [Google Scholar]
- 39. Lloyd D, Kainz B, van Amerom JF, et al Prenatal MRI visualisation of the aortic arch and fetal vasculature using motion‐corrected slice‐to‐volume reconstruction. J Cardiovasc Magn Reson 2016;18(Supp 1):P180. [Google Scholar]
- 40. Seed M, van Amerom JFP, Yoo S‐j, et al Feasibility of quantification of the distribution of blood flow in the normal human fetal circulation using CMR: a cross‐sectional study. J Cardiovasc Magn Reson 2012;14:79. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Al Nafisi B, van Amerom JF, Forsey J, et al Fetal circulation in left‐sided congenital heart disease measured by cardiovascular magnetic resonance: a case‐control study. J Cardiovasc Magn Reson 2013;15:65. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Borik S, Macgowan CK, Seed M. Maternal hyperoxygenation and foetal cardiac MRI in the assessment of the borderline left ventricle. Cardiol Young 2014;1–4:. [DOI] [PubMed] [Google Scholar]
- 43. Tsai‐Goodman B, Zhu MY, Al‐Rujaib M, et al Foetal blood flow measured using phase contrast cardiovascular magnetic resonance ‐ preliminary data comparing 1.5 T with 3.0 T. J Cardiovasc Magn Reson 2015;17:30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. van Amerom JF, Price AN, Alijabar P, et al Fetal cardiac cine imaging using realtime MRI and image‐based spatiotemporal motion‐correction. ESMRMB 2015, 32nd Annual Scientific Meeting, Edinburgh, UK, 1‐3 October: Abstracts, Thursday. Magnetic Resonance Materials in Physics. Biology and Medicine 2015;28:1–135. [Google Scholar]
- 45. Andescavage N, Yarish A, Donofrio M, et al 3‐D volumetric MRI evaluation of the placenta in fetuses with complex congenital heart disease. Placenta 2015;36:1024–30. [DOI] [PMC free article] [PubMed] [Google Scholar]