Table 4.
DENSE sequences | 1.5T | 3.0T | ||
---|---|---|---|---|
Image quality | Ecc (n = 81) | Ell (n = 34) | Ecc (n = 88) | Ell (n = 86) |
High qualitya | 76 (94%) | 31 (92%) | 34 (39%) | 28 (33%) |
Adequate qualityb | 5 (6%) | 3 (8%) | 48 (54%) | 49 (57%) |
Nondiagnosticc | 0 (0%) | 0 (0%) | 6 (7%) | 9 (10%) |
Number of images with motion artifact | 2 (2%) | 2 (6%) | 40 (45%) | 43 (50%) |
Number of images with field effect artifact | 1 (1%) | 0 (0%) | 11 (13%) | 13 (15%) |
High quality: Well defined endo‐ and epicardial borders at end‐systole. No ghosting due to patient breathing. No flow artifact.
Adequate quality: One or more of the following were present: Slight blurring of endo‐ and epicardial borders at end‐systole. Slight artifact due to flow within blood pool but not affecting myocardium. Slight ghosting due to patient breathing.
Nondiagnostic: One or more of the following are present: Loss of endo‐ and epicardial borders at end‐systole; ghosting due to respiration.