Skip to main content
. 2016 Sep 24;17(10):1627. doi: 10.3390/ijms17101627

Table 2.

Comparison between calibration functions, determination coefficients, LODs, LOQs and lineal dynamic ranges for the FLD and the MS detectors.

Analyte Detector External Calibration Curve R2 LOD (μg·mL−1) LOQ (μg·mL−1) Linear Range (μg·mL−1) a
Gal FLD y = 0.127x + 0.1263 0.9987 0.625 2.083 250
MS y = 40519x + 37647 0.9965 0.051 0.171 100
HTY FLD y = 10.226x + 2.1124 0.9989 0.035 0.115 20
MS y = 52263x + 55720 0.9945 0.017 0.057 50
TY FLD y = 6.239x + 3.3395 0.9993 0.009 0.029 100
MS y = 20379x + 17714 0.9924 0.058 0.195 50
4-HBA FLD y = 0.880x + 2.804 0.9954 0.036 0.121 250
MS y = 16457x + 49330 0.9911 0.061 0.204 150
4-HPA FLD y = 2.012x + 2.6778 0.9982 0.357 1.190 250
MS y = 13242x − 2729.7 0.9938 0.122 0.407 150
Van FLD y = 14.276x − 0.4684 0.9989 0.004 0.013 20
MS y = 16919x + 11618 0.9956 0.025 0.084 10
Syr FLD y = 6.770x + 8.7496 0.9861 0.007 0.023 50
MS y = 27088x + 14618 0.9928 0.025 0.083 10
Hmvan FLD y = 1.244x + 0.4454 0.9981 0.556 1.852 250
MS y = 31576x − 42195 0.9983 0.155 0.515 50
p-Cou FLD y = 0.272x + 1.036 0.9932 0.031 0.103 250
MS y = 40281x + 21180 0.9941 0.018 0.059 20
Val FLD y = 0.299x + 0.631 0.9974 0.090 0.301 250
MS y = 12440x + 7246.2 0.9927 0.055 0.182 10
Sin FLD y = 0.618x + 0.8065 0.9986 0.039 0.132 250
MS y = 58259x + 28312 0.9939 0.021 0.069 10
Fer FLD y = 1.287x + 3.1028 0.9965 0.021 0.069 250
MS y = 39342x + 13674 0.9949 0.022 0.073 10
m-Cou FLD y = 0.083x + 0.122 0.9995 0.306 1.020 250
MS y = 58990x + 130287 0.9934 0.013 0.043 100
Ole FLD y = 1.246x + 0.1575 0.9998 0.273 0.909 250
MS y = 6709x + 86 0.9990 0.050 0.165 20
o-Cou FLD y = 1.170x + 3.713 0.9954 0.022 0.072 250
MS y = 33291x + 12614 0.9946 0.077 0.255 100
Lut FLD y = 0.009x + 0.0064 0.9984 7.143 23.810 250
MS y = 201100x + 44947 0.9994 0.005 0.016 10
Pin FLD y = 40.452x + 7.795 0.9991 0.005 0.016 10
MS y = 34634x + 63107 0.9912 0.032 0.108 50

a Linear ranges were established from LOQ to the indicated value.