Table 3.
Study | Patients | Prosthesis | Followup (months) | Recurrence rate | Flexion/extension | Supination/pronation | Complications | Functional scores |
---|---|---|---|---|---|---|---|---|
Gold [12] 1957 | 1 | Distal radius prosthesis | 59 | 0 | NA | NA | The fracture of the metallic stem | NA |
Hariri et al. [14] 2013 | 1 | Mega wrist prosthesis | 33 | 0 | 20°/70 | 60°/70° | Dorsal subluxation of the ulnar head | QuickDASH score, 52.27/100 Enneking score, 83% |
Hatano et al. [15] 2006 | 2 | Distal radius ceramic prosthesis | 168, 120 | 0 | 15°/30° 15°/35° |
45°/30° 50°/60° |
DJD (2), radiolucent line of bone-cement interface (2) | Enneking score, 83% Enneking score, 83% |
Natarajan et al. [25] 2009 | 24* | Custom-made bipolar hinge prosthesis | 78 (range, 24–156) | 0 | 25° (range, 5°–40°)/20° (range 10°–50°) | 40° (range, 5°–80°)/60° (range, 25°–85°) | Skin flap necrosis (2), infection (2), aseptic loosening (2) | Averaged MSTS score, 75% |
Gokaraju et al. [11] 2009 | 1 | Predominantly titanium prosthesis | 56 | 0 | 20°/40° | Pronation full while supination 45° | Mild degenerative changes of the carpal bones | Full DASH score, 10.3/100 |
Zhang et al. [35] 2015 | 11 | Custom-made distal radius prosthesis | 55.5 (range, 24–83) | 9.09 | 30° (range, 15°–45°)/40.9° (range, 20°–60°) | 46.4° (range, 20°–65°)/38.2° (range, 10°–60°) | Superficial infection (1), pain (1) | MSTS score, 80 (range, 63.3–93.3) |
Current study | 10 | Custom-made distal radius prosthesis | 52 (range, 24–90) | 0 | 20°(range, 10°–38°)/22°(range, 15°–39°) | 38° (range, 15°–65°)/36° (range, 22°–60°) | Pain (2), DJD (3), aseptic loosening (1), wrist subluxation (2) | Modified Mayo wrist score, 68 (range, 45–90) |
GCT = giant cell tumor; * of 24 patients, 16 had giant cell tumors; DJD = degenerative joint disease; NA = not applicable; MSTS = Musculoskeletal Tumor Society.