Skip to main content
. Author manuscript; available in PMC: 2017 Nov 1.
Published in final edited form as: J Photochem Photobiol B. 2016 Sep 12;164:96–102. doi: 10.1016/j.jphotobiol.2016.09.017

Table 1.

Laser Parameters

Irradiation and Treatment Parameters. The parameters were obtained at the IBRAMED (Brazil) for the AsGa laser, 904nm; model Laserpulse, handheld probe, used in contact to the mice skin.

Parameter [unit] value
Center wavelength [nm] 904
Operating mode pulsed
Frequency [Hz] 9500
Pulse duration [ns] 60
Duty cycle [%] 20
Energy per pulse [J] 42 nJ
Peak power [W] 70
Average power [mW] 39.9
Polarization No
Spot size [cm2] 0.1309
Beam shape elliptical
Beam profile Gaussian
Irradiance at target [mW/cm2] 304.8
Exposure duration [sec] 60
Radiant exposure [J/cm2] 18.288
Total Radiant energy [J] 2.394
Number of points irradiated 1
Area irradiated [cm2] 0.1309
Application technique Contact
Number and frequency of treatment sessions 1 × day / 5 days