Table 1.

BE% (Equation (6)) for the biosensor designs PtC/PoPD/PEI2/GluOx5/XL fabricated using a range of crosslinkers (XLs) at different concentrations and crosslinking conditions determined on day 0 and day 10. ΔBE% (day 0) is the difference in the day-0 BE% value compared with the crosslinker-free design. The 10-day stability value was calculated as the change in BE% on day 10 compared with that of day 0 for each biosensor design. The results for PEGDE crosslinked at room temperature or at 55 °C were statistically the same, and were, therefore, pooled.

Crosslinker BE% (Day 0) BE% (Day 10) ΔBE% (Day 0) p-Value 10-Day Stability p-Value
none 41 ± 2 (n = 16) 26 ± 3 (n = 14) 0 N/A −15 ± 4 <0.001
GA (1%) 37 ± 1 (n = 2) 23 ± 2 (n = 2) −4 ± 2 >0.61 −14 ± 3 <0.03
GA (25%) 39 ± 3 (n = 8) 45 ± 3 (n = 6) −3 ± 4 >0.57 +6 ± 4 >0.19
PEG (0.1%) 45 ± 2 (n = 3) 19 ± 6 (n = 3) +4 ± 3 >0.41 −26 ± 5 <0.02
PEG (1%) 45 ± 8 (n = 3) 21 ± 6 (n = 3) +4 ± 7 >0.46 −24 ± 9 <0.07
PEGDE (0.1%) 39 ± 2 (n = 7) 34 ± 3 (n = 7) −2 ± 3 >0.55 −5 ± 4 >0.28
PEGDE (1%) 34 ± 2 (n = 12) 35 ± 3 (n = 12) −7 ± 3 <0.05 +1 ± 4 >0.78