Table 3.

Comparison of state of the art works. SPS, samples per second; ESB, Enhanced ShockBurst.

This Work Neurochip-2 WIMAGINE PennBMBI W-HERBS Nguyen Angotzi
(2015) (2011) [9] (2015) [8] (2015) [11] (2011) [7] (2014) [10] (2014) [12]
Recording channel 32 unipolar/bipolar 3 unipolar/bipolar 64 4 64×2 32 8
Gain 192 1000 or 5000 1000 200 40–80 dB 200 -
ADC resolution 16 bits 8 bits 12 bits 12bits 12 bits 16 bits 8 bits
Sampling rate 800 SPS/channel up to 30 kSPS/channel 256 SPS 1 kSPS 21 kSPS 1 kSPS 400 kSPS
12.5 kSPS/channel
15 kSPS/channel
Bandwidth 0.1–20 kHz 10 Hz–7.5 Hz 0.5 Hz–400 Hz 0.05 Hz–6 kHz 0.1 Hz–1 kHz 0.2 Hz–5 kHz 1 Hz–10 kHz
Communication 2.4 GHz RF communication link with μESB protocol Serial cable/ infrared data link Proprietary UHF link in MICS band 2.4 GHz RF communication link Bluetooth Wired 2.4 GHz ISM band
Power supply Rechargeable 3.7 V battery 1 or 2 rechargeable 3.6 V battery Inductive link Rechargeable 3.7 V battery Polymer Lithium ion 3.7 V battery 5 V USB 3.7 V 700 mAh battery
Power consumption 4.22–15.4 mA 284–420 mW 75 mW w/o charging
350 mW w/ charging
7.3 mA transmit for sensor node only 4.9 mW (AFE)
300 mW (wireless)
- -
Battery charging Wireless ultrasonic charging - - - Wireless charging coil 4 W at distance 38 mm - -
Size 35 mm in diameter
10 mm in thickness
63×63×30 mm 50 mm in diameter
12.54 mm in thickness
antenna:10 cm2
56×36×13 mm + 43×27×8 mm + 31×13×8 mm 20×30×2.5 mm + 60×60×8 mm + 40 mm in diameter
8 mm in thickness
29.5×43.4 mm -
Stimulation channel 4 unipolar/bipolar 3 unipolar/bipolar - 2 - 1 optical stimuli 8 bipolar
Stimulation voltage 20 V ±15 V (normal)
±50 V (high V)
- ±12 V - - ±9 V
current intensity 40 μA 10–200 μA current pulse or 0.5–5 mA - 0–1 mA - - 300 μA
Pulse width 1 us minimum 0.2 ms, 0.6 ms - 200 μs - - -
Pulse frequency 250 Hz 1/min or 1/10 min - - - - -