Skip to main content
. 2016 Sep 20;105(11):e499–e505. doi: 10.1111/apa.13553

Table 4.

False‐positive rate of screening methods for detecting Critical CHD between different gestational ages in NICU

POX Clinical evaluation POX + clinical evaluation
≤35 weeks >35 weeks ≤35 weeks >35 weeks ≤35 weeks >35 weeks
True positives 6 10 3 9 7 12
False negatives 1 2 4 3 0 0
False positives 1384 908 483 468 1454 1012
True negatives 928 889 1829 1329 858 785
False‐positive rate (%) 59.9 50.5 20.9 26.0 62.9 56.3
Sensitivity (%) 85.7
(48.7, 97.4)
83.3
(55.2, 95.3)
42.9
(15.8, 75.0)
75.0
(46.8, 91.1)
100
(64.6, 100)
100
(75.8, 100)
Specificity (%) 40.1
(38.2, 42.2)
49.5
(47.2, 51.8)
79.1
(77.4, 80.7)
74.0
(71.9, 75.9)
37.1
(35.2, 39.1)
43.7
(41.4, 46.0)

Data: number or percentage (95% CI).