Published in final edited form as: Eur J Nucl Med Mol Imaging. 2016 Aug 25;43(13):2374–2380. doi: 10.1007/s00259-016-3477-3

Table 2.

Receiver operating characteristic analyses of sensor diagnostic performance. Ability of integration metrics from the sensor TACs to predict whether an infiltration will be visible on ensuing PET: infiltration visible (n = 4); infiltration not visible (n = 6).

TAC interval Sensor arm ROC area Sensitivity Specificity Cut-off value PPV NPV
Peak only Injection 0.71 50% 100% 1.76 × 105 100% 75.0%
Peak only I/C ratio 0.58 50% 100% 29.20 100% 75.0%
0–10 min Injection 0.92 75% 100% 3.34 × 105 100% 85.7%
0–10 min I/C ratio 0.92 75% 100% 2.62 100% 85.7%
0–20 min Injection 0.92 75% 100% 5.40 × 105 100% 85.7%
0–20 min I/C ratio 0.92 75% 100% 1.87 100% 85.7%
0–30 min Injection 0.88 75% 100% 7.48 × 105 100% 85.7%
0–30 min I/C ratio 0.88 75% 100% 1.61 100% 85.7%
0–40 min Injection 0.83 75% 100% 9.45 × 105 100% 85.7%
0–40 min I/C ratio 0.83 75% 100% 1.47 100% 85.7%
30–40 min Injection 0.79 75% 100% 1.97 × 105 100% 85.7%
30–40 min I/C ratio 0.88 75% 100% 1.31 100% 85.7%

TAC, time activity curve; ROC, receiver operating characteristic; I/C, injection arm/control arm; PPV, positive predictive value; NPV, negative predictive value.