Figure 4.
a, assay repeatability. In patient subset duplicate blood tubes were collected at same time point for repeatability analysis. Normalized CTC counts are shown on log scale with reference to low/medium/high classification. Concordance among these 15 matched samples using this scheme was 100%. b, comparison to CellSearch. First 5 patients had matched samples sent for analysis on IsoFlux System and CellSearch CTC Test at baseline and after 1 chemotherapy cycle. In 5 patients tested no cells were detected using CellSearch.