Published in final edited form as: Optom Vis Sci. 2016 Nov;93(11):1356–1363. doi: 10.1097/OPX.0000000000000987

Table 2.

Comparison of repeatability values with previous studies of simulated keratometry.

Keratometry Value No of eyes Study Instrument COR COV ICC
Steep K (D) 264 current study Zeiss Atlas 0.36 0.30 0.99
79 Huang et al (2015)* OphthaTop 0.29 0.24 0.99
35 Wang et al (2012)* EyeSys Vista 0.35 0.29 0.99
Medmont 0.22 0.17 0.99
Topolyzer 0.34 0.28 0.99
77 Mao (2013)* Keratograph 0.30 0.24 0.99
57 Kobashi (2012) Zeiss Atlas 0.98

Flat K orientation (Degrees) 140 current study Zeiss Atlas 8.75 4.71 0.99
57 Kobashi (2012) Zeiss Atlas 0.98

Astigmatism(D) 264 current study Zeiss Atlas 0.29 10.53 0.97
57 Kobashi (2012) Zeiss Atlas 0.96

J0 (D) 264 current study Zeiss Atlas 0.15 0.98
79 Huang et al (2015)* OphthaTop 0.16 0.97
77 Mao (2013)* Keratograph 0.09 0.99
57 Kobashi (2012) Zeiss Atlas 0.92
101 Read (2009) Medmont 0.12 0.99

J45 (D) 264 current study Zeiss Atlas 0.17 0.90
79 Huang et al (2015)* OphthaTop 0.12 0.92
77 Mao (2013)* Keratograph 0.08 0.98
57 Kobashi (2012) Zeiss Atlas 0.91
101 Read (2009) Medmont 0.12 0.95
*

= values are average of 2 observers;

COR= coefficient of repeatability; COV = coefficient of variation and ICC = Intraclass correlation coefficient