Table 1.

Compliance, kp (m2Pa−1), static pulse wave velocity estimate, cstatic (ms−1), and the effective elastic modulus, Eeff(kPa), measured using static testing for each phantom. Values are given at a diameter of 8.7mm, which is approximately the diameter in the phantom PWI experiments. Eeff is the elastic modulus of a tube of the average phantom thickness (3.64mm) which gives the appropriate kp

Material kp ×108 cstatic Eeff
P1-Soft 0.96±0.18 2.49 11.1
P1-Stiff 0.51±0.06 3.41 20.9
P2-Soft 1.04±0.15 2.39 10.2
P2-Stiff 0.64±0.11 3.06 16.7
P3-Soft 1.14±0.07 2.28 9.35
P3-Stiff 0.65±0.04 3.02 16.4
P4-Soft 1.23±0.07 2.20 8.69
P4-Stiff 0.79±0.09 2.74 13.4