. 2016 Nov 19;28(1):1. doi: 10.1007/s10856-016-5790-6

Table 1.

Adhesion testing of different compositions of the reactive isocyanate-terminated hyper-branched adhesives using a bucket-handle meniscus tear model

Glue Maximum adhesive force [N] Displacement at maximum force [mm] Energy required to open glued tear [mJ] Energy required to detach adhesive [mJ]
Hyper-branched adhesive 3.7 ± 0.97 2.4 ± 0.96 11.0 ± 5.2 6.1 ± 3.5
Hyper-branched adhesive + spermidine 5:1 3.7 ± 1.5 1.6 ± 0.4 5.5 ± 3.7 3.2 ± 2.1
Hyper-branched adhesive + DABCO 3.2 ± 0.47 2.8 ± 0.6 10.3 ± 2.1 4.8 ± 1.1
Fibrin glue 0.3 ± 0.04* 6.0* 1.6 ± 0.1** 0

Data are presented as mean ± standard deviation

* The value is the force at the maximum displacement of 6 mm, a maximum force was not reached

** These values are the areas under the curve for displacement from 0 to the maximum displacement of 6 mm