Table 4.

Liver Iron Quantification

Description T2w
Coronal
T1w In/
Out
T2w
FatSat
DWI b= 0, 50,
500
T1w
Axial
T2*MAP In-Phase T2*MAP Short
ET
T2*MAP Long ET
Sequence SSFSE 2D SPGR FRFSE EPI 3D SPGR 2D GRE 2D GRE 2D GRE
Plane Coronal Axial Axial Axial Axial Axial Axial Axial
Slice thickness
(mm)
5 7 7 7 5 10 (3 slices) 10 (3 slices) 10 (3 slices)
Gap (mm) 0 1 1 1 −2.5 2 2 2
Fat suppression Spectr. (Spectr.) IR
TE1/TE2 (ms)
(out/in)
100 2.2/4.4 102 Min (<90) ? 1.9; 4.7; 7.5; 10.2; 13; 15.8;
18.5; 21.3
0.9; 1.8; 2.8; 3.8;
4.8; 5.8
0.9; 2.0; 3.1; 4.2; 5.3; 6.4; 7.6; 8.7; 9.8;
10.9; 12; 13.1
TR (ms) 210 2500 4000 (2000) ? −40 35.5 35.5
Flip angle 90 90 90 90 12 25 12 12
ETL 15
NEX 0.5 1 1 1 (12) <1 1 1 1
Phase enc.
steps
320 256 256 128 256 384 128 104
Frequency steps 192 128 160 128 192 192 64 104
Breath BH BH BH BH/Trig. BH BH BH BH
Notes 8 echoes 6 echoes 12 echoes

After basic sequences for evaluation of liver parenchyma, 3 multiecho T2*-weighted sequences are acquired to obtain good accuracy in estimation. In the first T2*-weighted series, ETs are set approximately within water and fat phase coherence, whereas in short ET the first echo time is <1 ms to investigate higher iron concentrations. The long ET series is designed to investigate lower iron concentrations. BH, breath-hold; Dixon, Dixon technique; enc., encoding; ETL, echo train length; FSE, fast spin echo; In/out, in-phase and out-of-phase; IR, inversion recovery; NEX, number of excitations; Spectr., chemically selective fat suppression; SSFSE, half Fourier single-shot FSE. Trig., respiratory triggering; w, weighted.