Table 2.
Differences in IOP between different tonometers and Goldmann applanation tonometry [3].
Tonometer | Mean difference between tonometer and GAT | 95% confidence interval | 95% limits of agreement | % within 2 mmHg | ||
---|---|---|---|---|---|---|
DCT | 1.8 | +1.3 | +2.3 | −3.0 | +6.6 | 47 |
NCT | 0.3 | −0.1 | +0.7 | −3.5 | +4.0 | 69 |
ORA | 1.5 | +0.9 | +2.2 | −4.3 | +7.3 | 45 |
Ocuton S | 2.7 | −1.2 | +6.7 | −4.0 | +9.6 | 33 |
RT-(Icare) | 0.9 | +0.5 | +1.5 | −4.3 | +6.3 | 51 |
TonoPen | 0.2 | −0.4 | +0.9 | −5.2 | +5.7 | 52 |
Transpalpebral | −0.5 | −1.3 | +0.3 | −7.0 | +5.9 | 45 |
DCT, dynamic contour tonometer; NCT, noncontact tonometer; ORA, ocular response analyzer; RT, rebound tonometer.
Reproduced from [3] with permission from the European Glaucoma Society.