Table 1.

Thermal stress and phase transition of anti CD40 antibody as a function of heating rate (SE corresponds to fit error).

Instrumentation Buffer pH (Initial) Heating Rate (°C/hr) Tm ± S.E. (°C)
Fluorescence PBS 7.0 60 64.5 ± 0.6
CD PBS 7.0 30 68.3 ± 0.4
CD PBS 7.0 60 69.1 ± 0.3
CD PBS 7.0 120 70.5 ± 0.8
CD Acetate 4.0 30 68.1 ± 1.8
CD Acetate 4.0 60 69.1 ± 1.4
CD Acetate 4.0 120 70.4 ± 2.0