Skip to main content
. 2016 Sep 22;46(1):105–109. doi: 10.1007/s00256-016-2478-4

Fig. 6.

Fig. 6

Photo of the revised Adept hip resurfacing removed from the 53-year-old gentleman. Manufacturer labels in the surgical notes record the components as being a 56-mm cup designed for a 48-mm head and a 48-mm head. Engravings (red arrows) and coordinate measuring machine measurements confirm the CSM, as the cup is actually a 56-mm cup designed for a 50-mm head