Table 1.
Area scanned | kV | mAs | Scan length (cm)a | Collimation |
---|---|---|---|---|
ASIS–stem tip | 100 | 100 | 30 | 128*0.625 |
Knee | 100 | 70 | 20 | 128*0.625 |
Ankle | 100 | 45 | 10 | 24*1.25 |
ASIS anterior superior iliac spine
aScan length varies depending upon patient size and component. An anterior–posterior topogram should be acquired to determine scan range. All areas should be scanned using a pitch of 1.0 and rotation time of 1.0 s and reconstructed using 1-mm slices and the kernel “BONE”. An extended scale should be used.