Skip to main content
. 2016 Jul 11;89(1064):20151025. doi: 10.1259/bjr.20151025

Table 2.

MR imaging and MR cholangiography parameters and acquisition protocol

MR sequence T2 weighted 2D TSE T1 weighted 2D GRE T2 weighted 3D ISO “respiratory triggered” T2 weighted 2D SS TSE T1 weighted 3D GREa
Fat suppression With/without Without With With With
Contrast agent Pre-contrast images assessed Pre-contrast images assessed Pre-contrast images assessed Pre-contrast images assessed Pre- and post-contrast images assessed
TR/TE 4000/176 140/2.2–4.4 2500/683 4500/754 5.7/2.8
Flip angle (degrees) 150 90 140 180 10
Section thickness (mm) 5 5 1 50 2.5
Matrix 192 × 256 192 × 256 384 × 341 384 × 341 192 × 256
Bandwidth (Hz/pixel) 260 260 150 150 250
Field of view (cm) 30–50 30–50 30–50 30–50 30–50

2D, two-dimensional; 3D, three-dimensional; GRE, gradient recalled echo; ISO, isotropic; SS TSE, single shot turbo spin echo; TE, echo time; TR, repetition time; TSE, turbo spin echo.

a

T1 weighted 3D spoiled gradient recalled echo images were acquired at approximately 25, 60 and 150 s after contrast administration during the arterial, portal venous and delayed phases, respectively, and during the delayed hepatobiliary phase at approximately 20 min (gadoxetic acid ) or 75 min (gadobenate dimeglumine).