Skip to main content
. 2016 Aug 9;89(1065):20160512. doi: 10.1259/bjr.20160512

Table 3.

Apparent diffusion coefficient (ADC) values obtained with single-shot echo-planar imaging diffusion-weighted imaging (EPI-DWI) and single-shot turbo spin echo diffusion-weighted imaging (TSE-DWI)

Acquisition techniques ADC values (×10−3 mm2 s−1)
BG Comp. 1 Comp. 2 Comp. 3 Comp. 4
TSE-DWI
 AF1 1.97 ± 0.008 1.68 ± 0.005 1.31 ± 0.008 0.932 ± 0.004 0.700 ± 0.004
 AF2 1.85 ± 0.004 1.59 ± 0.005 1.32 ± 0.004 0.943 ± 0.004 0.700 ± 0.004
 AF3 1.82 ± 0.009 1.58 ± 0.005 1.29 ± 0.004 0.903 ± 0.003 0.638 ± 0.003
 AF4 1.75 ± 0.015 1.59 ± 0.006 1.29 ± 0.009 0.983 ± 0.005 0.753 ± 0.007
EPI-DWI 1.92 ± 0.005 1.70 ± 0.005 1.43 ± 0.005 1.04 ± 0.015 0.881 ± 0.014

AF, acceleration factor; BG, background; Comp., compartment.

ADC values are presented as mean ± standard deviation.