Skip to main content
. 2016 Nov 15;8(11):4523–4531.

Table 2.

Comparison of the three categories of dual-energy CT

Dual-Source system Single-Source System Dual-Layer Detector system
X-ray tube Two One One
Detector Two One Two
Tube potential switch - During each scan (0.2 ms) Between each scan (<1 s) -
Voltage 140 kV (tube A) 140 kV and 80 kV 135 kV and 80 kV 80/120/140 kV
80 kV (tube B)
Temporal resolution 66 ms 165 ms 177 ms 127 ms
Spatial resolution 0.40 mm3 0.23 mm3 0.50 mm3 0.625 mm3
Scanning time 0.25 s 0.35 s 0.27 s
Heart scanning time 0.25 s 7 s 1-2 s 2 s
Scanning mode High-spiral acquisition Retrospective ECG-gated spiral acquisition
Or shuttle mode
Advantage High temporal resolution Exact BH correction Better material decomposition The two views exactly registered
Disadvantage Degraded CNR Imaging artifacts Motion artifacts Suboptimal material decomposition
Data truncation artifacts BH artifacts
Example of representing vendor Siemens GE Toshiba Philips

ECG, Electrocardiograph; BH, Beam hardening; CNR, Contrast to noise ratio.