Overview of the HC2, cobas and Aptima test specifications and accompanying automated testing solutions

Assay Nr. of devices HrHPV types detected Target gene Detection technology Input volume Dead volume Control
HC2 4 16/18/31/33/35/39/45/51/52/56/58/59/68 Entire hrHPV genome (DNA) Nucleic acid hybridization assay with signal amplification 3.0 mL 5.0 mL Process control for each sample batch
cobas 3 16/18/31/33/35/39/45/51/52/56/58/59/66/68 L1 (DNA) Real time Polymerase Chain Reaction 0.25 mL 0.5 mL ß-globin for process validity of individual sample and input sufficiency
Aptima 2 16/18/31/33/35/39/45/51/52/56/58/59/66/68 E6/E7 (RNA) Transcription Mediated Amplification 0.4 mL 5.0 mL Spiked internal control for process validity of individual sample

HC2 Hybrid Capture 2