Table 1. Guide to modifying the DIA method.

Goal Selectivity m/z Range
Covered
Number of
Sample
Injections
Chromatographic
Sampling Rate
Increase sensitivity
  Reduce isolation width
  Reduce isolation width and
use multiple injections per
sample
  Increase max ion inject time*
  Increase resolving power and
max ion inject time*
↓↓
Sample more peptides
  Use multiple injections per
sample
  Increase isolation width
Improve chromatogram
sampling
  Increase isolation width
  Reduce m/z range covered
*

Instruments with an ion trap mass analyzer that use automatic gain control (e.g. Thermo Scientific LTQ, Velos, Orbitrap, and Exactive series).