Published in final edited form as: Methods Enzymol. 2016 Jun 16;579:51–86. doi: 10.1016/bs.mie.2016.04.011

Table 2.

Specimen support geometries for particular applications

Mode Grid Foil Film
Negative stain 400 mesh none 50–100 Å am-C
Diagnostic cryo 300 mesh 1.2/1.3 μm none / 20 Å am-C
Medium-resolution (≥3.5 Å) cryo 300 mesh 1.2/1.3 μm none / 20 Å am-C
High-resolution cryo (<3.5 Å) >400 kDa 300 mesh 1.2/1.3 μm none / 20 Å am-C
High-resolution cryo (<3.5 Å) <400 kDa 300 mesh 1.2/1.3 μm none / graphene
Very high-resolution cryo (<2.8 Å) 300 mesh 0.6/1.0 μm none / graphene
Cryo-tomography: cellular (>30 Å) 200 mesh 2.0/2.0 μm * none
Cryo-tomography: high-resolution sub-tomogram avg. (<15 Å) 300 mesh 1.2/1.3 μm none / 20 Å am-C
*

Larger holes may be used for particularly large specimens.