. 2016 Nov 23;2016:5075182. doi: 10.1155/2016/5075182

Table 2.

Horizontal distortion of 3D images assessed by mean absolute error in arch widths between digital and actual measurements (unit: mm).

E4D Dentist iTero
Model B Model L Significant Model B Model L Significant
Maxillary intermolar width 1.76 (1.7) 1.23 (1.32) NS 0.36 (0.31) 0.29 (0.32) NS
Mandibular intermolar width 0.44 (0.44) 0.29 (0.31) NS 0.18 (0.39) 0.60 (0.16)
Maxillary intercanine width 0.93 (0.59) 0.56 (0.82) NS 0.18 (0.15) 0.34 (0.10)
Mandibular intercanine width 0.39 (0.26) 0.17 (0.24) NS 0.17 (0.27) 0.49 (0.56)

Trios Zfx IntraScan
Model B Model L Significant Model B Model L Significant

Maxillary intermolar width 0.42 (0.58) 0.26 (0.34) NS 0.79 (1.07) 0.69 (0.95) NS
Mandibular intermolar width 0.15 (0.11) 0.38 (0.44) 0.60 (1.06) 0.81 (1.03) NS
Maxillary intercanine width 0.20 (0.16) 0.18 (0.16) NS 0.09 (0.27) 0.26 (0.31) NS
Mandibular intercanine width 0.35 (0.21) 0.46 (0.22) NS 0.17 (0.40) 0.45 (1.44) NS

( ): maximum deviation, which is the difference between maximum and minimum errors of 5 repeated images, which refers to the largest distortion considered as precision of the scanner.

P < 0.05 (Mann–Whitney U test).

NS: not significant.