Table 1.

Discrepancies in position detection between ExacTrac and CBCT (N = 42)

mean ± SD median range 95% (abs.)
lateral +0.1 ± 0.3 +0.1 –0.9 to +0.8 <0.8
longitudinal –0.5 ± 0.4 –0.4 –1.6 to +0.2 <1.2
vertical –0.3 ± 0.5 –0.4 –1.4 to +0.8 <1.0
radial 0.8 ± 0.3 0.8 0.3 to 1.8 <1.4

CBCT Cone Beam Computed Tomography

All units are in mm