. 2016 Dec 7;11:158. doi: 10.1186/s13014-016-0735-2

Table 4.

Shifts of a head phantom as preset by the robotic couch and as detected by both ExacTrac and CBCT (N = 24, resp. 8 per axis)

Discrepancy between … … ExacTrac and robotic couch … CBCT and robotic couch … ExacTrac and CBCT
 lateral 0.0 ± 0.5 (p = 0.9) 0.0 ± 0.6 (p = 1.0) 0.0 ± 0.1 (p = 0.5)
 longitudinal 0.4 ± 0.3 (p = 0.1) 0.3 ± 0.3 (p = 0.1) 0.1 ± 0.1 (p = 0.3)
 vertical 0.0 ± 0.4 (p = 0.9) 0.0 ± 0.4 (p = 1.0) 0.0 ± 0.1 (p = 0.5)
 radial 0.7 ± 0.3 (p = 0.007) 0.7 ± 0.3 (p = 0.006) 0.1 ± 0.0 (p < 0.001)

All units are in mm, average ± SD (p values from two-sided t-test)