Fitted parameters corresponding to the different HIFU pulses: the instant of maximum temperature (t max), maximum temperature (°Cmax), and the calculated equivalent thermal dose CEM43. C and D are the Parker’s fitted constants (see Additional file 2)

Label tmax (s) Tmax (°C) CEM43 C D
58 °C 13.01 57.6 1503.0 3.002 0.168
55 °C 13.14 55.0 199.6 2.287 0.127
HT 8.08 53.7 60.0 3.002 0.127