Table 6.

Recommended limits for viable airborne particles in the environment according to FDA-cGMP, EU-GMP, and USP.

Clean area type Maximal number of cfu in the environment
FDA-cGMP EU-GMP USP
FS209E ISO EU-GMP Air sample (cfu/m3) Settle platesa (diam. 90 mm; cfu/4 h) Air sample (cfu/m3) Settle plates (diam. 90 mm; cfu/4 h)b Air sample (cfu/m3)
M3.5 (100) 5 A 1c 1c <1 <1 <3
M4.5 (1,000) 6 7 3
M5.5 (10,000) 7 B 10 5 10 5 <20
M6.5 (100,000) 8 C 100 50 100 50 <100
D 200 100

aThe additional use of settling plates is optional.

bIndividual settle plates may be exposed for less than 4 hours.

cSamples from class 100 (ISO 5) environments should normally yield no microbiological contaminants.