TABLE I.

Protocol information for the Elekta XVI, Varian OBI, and GE LightSpeed system scans used in this study.

Protocol Collimator/filter settings Field of view (cm) Matrix size Pixel size (mm) Slice thickness (mm) Tube voltage (kVp) Total mAs Frames Scan angle (deg)
Elekta XVI Pelvis M20 M20/F1 42.64 410 × 410 1.0 2.5 120 1056 660 360
Elekta XVI Prostate Seed S10 S10/F0 27.68 270 × 270 1.0 2.5 120 117.1 366 200
Elekta XVI Chest M20 M20/F1 42.64 410 × 410 1.0 2.5 120 1056 660 360
Elekta XVI Head and Neck S20 S20/F0 27.68 270 × 270 1.0 2.5 100 36.6 366 200
Varian OBI Pelvis Half-Fan 46.0 512 × 512 0.90 2.5 125 1056 660 364
Varian OBI Spotlight Full-Fan 26.0 512 × 512 0.51 2.5 125 1320 367 200
Varian OBI Thorax Half-Fan 46.0 512 × 512 0.90 2.5 125 264 660 364
Varian OBI Head Full-Fan 26.0 512 × 512 0.51 2.5 100 264 367 200
GE LightSpeed Pelvis (Helical Mode) 46.0 512 × 512 0.90 2.5 120 1056 660
GE LightSpeed Head (Helical Mode) 26.0 512 × 512 0.51 2.5 120 117.1 366